Temperature Calibrator

Temperature Calibrator

Temperature Calibrator

Accuracy ± 0.05%

Output Range Output Range Resolution Accuracy Remarks
Voltage 100mV -10.00~110.00mV 0.01mV ±0.05% Setting value ± 30uV The max. Output
current ± 5mA
1000mV -100.00~1100.00mV 0.1mV ±0.05% Setting value ± 0.3mV
Resistance 400 ? 0.0-400.0 ? 0.1 ? ±0.05% Setting value ± 0.2? 1mA exciting current.
Thermo-
Couple
R -40~1760OC 1°C ±0.05% Setting value ± 3OC(<or=100OC) Adopt the ITS-90
temperature
standard
(Note 3)
S -20~1760OC 1°C ±0.05% Setting value ± 2OC(>100OC)
B -400~1800OC 1°C ±0.05% Setting value ± 3OC(400~600OC)
±0.05% Setting value ± 2OC(>600OC)
E -200.0~1000.0OC 0.1°C ±0.05% Setting value ± 2OC(<or=100OC)
±0.05% Setting value ± 1OC(>-100OC)
K -200.0~1370OC 0.1°C
J -200.0~1200.0OC 0.1°C
T -200.0~400.0OC 0.1°C
N -200.0~1300.0OC 0.1°C
Thermo-
resistance
Pt-100 -200.0~850.0OC 0.1°C ±0.05% Setting value ± 0.6OC Adopt the PT100-385
1mA (Note1,2)
Cu 50 -50.0~150.0OC 0.1°C

 

Note :

1. Not including the accesories lead resistance.
2. The range of exciting current : 0.5mA ~ 2mA, the max. Output voltage £ 2V.
3. Not including the accuracy of inner temperature compensatory transducer.
The range of inner temperature compensatory transducer : -10 - 50°C,
compensatory error ? 0.5°C
4. Temperature coefficient : ± 0.005%, range / °C (0°C- 18°C, 28°C - 50°C).